Temperature measurement in paediatrics

Posted: Jan 1 2000 Updated: Oct 15 2015

The Canadian Paediatric Society gives permission to print single copies of this document from our website. For permission to reprint or reproduce multiple copies, please see our copyright policy.

Principal author(s)

D Leduc, S Woods; Canadian Paediatric Society , Community Paediatrics Committee

Despite the fact that temperature measurement in children seems so simple – a wide variety of devices are available to record a fever from skin, oral or rectal mucosa or the tympanic membrane – the choice for health professionals and parents has never been so complicated.

According to traditional teaching, the normal body temperature is 37°C (98.6°F), but it is generally accepted that a temperature of 38°C (100°F) or greater, as measured by a rectal thermometer, represents a fever [1][2].

In febrile children younger than 36 months of age, most serious illnesses are caused by infectious agents [3]-[6]. The presence of a fever in children younger than three months of age triggers a thorough investigation into the source of the infection [7][8]. However, the presence of a normal or subnormal temperature in children younger than three months of age can also be associated with severe infections in the presence of other appropriate signs and symptoms. The definition of a fever of unknown origin also relies on stringent diagnostic criteria (ie, a fever lasting more than 14 days with no etiology found after routine tests), and depends on precise temperature recordings [9]-[11]. Finally, an appropriate recording of the absence of a fever reassures both parents and health care providers who seek to diminish fever phobia, and inappropriate medical consultations and investigations [12]. It is, therefore, essential that the measurement of a fever be accurate and reproducible from infancy through adolescence.

Current measurements and methods

Rectal thermometry

Rectal thermometry has traditionally been considered the gold standard for temperature measurement [13][14][39][44], but some studies have revealed limitations of this method [15]-[18]. Rectal temperatures are slow to change in relation to changing core temperature, and they have been shown to stay elevated well after the patient’s core temperature has begun to fall, and vice versa. Rectal readings are affected by the depth of a measurement, conditions affecting local blood flow and the presence of stool. Rectal perforation has been described [19][20], and without proper sterilization techniques, rectal thermometry has the capacity to spread contaminants that are commonly found in the rectum. Contraindications of rectal thermometry are addressed in the section Other Devices, below.

Many parents and older children may be uncomfortable with this method of temperature assessment, and older children may resent it.

Axillary thermometry

While axillary temperature is easy to measure (compared with oral or rectal measurements), it has been found to be an inaccurate estimate of core temperature in children [13][15][18][21]. This type of measurement relies on the thermometer remaining directly in place over the axillary artery, and it is largely influenced by environmental conditions.

Despite its low sensitivity and specificity in detecting fever, axillary temperature is recommended by the American Academy of Pediatrics as a screening test for fever in neonates because of the risk of rectal perforation with a rectal thermometer [22], although this complication is estimated to occur in less than 1 in 2 million measurements [23].

Oral thermometry

The sublingual site is easily accessible and reflects the temperature of the lingual arteries. However, oral temperature is easily influenced by the recent ingestion of food or drink and mouth breathing [21]. Oral thermometry relies on the mouth remaining sealed, with the tongue depressed for 3 to 4 min, which is a difficult task for young children. This method of temperature measurement cannot be used in young children, or in unconscious or uncooperative patients. Generally, it has been suggested that the accuracy of oral thermometry lies somewhere between that of axillary and rectal thermometry. It appears that accuracy may increase with the age of a child, primarily due to compliance and the ability to use proper technique.

Digital vs. mercury thermometers

The traditional mercury thermometer has been replaced by the more “user friendly” digital thermometer. Since the accuracy is comparable with both instruments [24] and mercury remains an environmental hazard, the Canadian Paediatric Society no longer recommends the use of mercury thermometers.

Tympanic thermometry

The first devices used to measure tympanic membrane (TM) temperature did so by being in direct contact with the tympanic membrane. In 1969, it was shown that such a device measured core temperature better than a rectal thermometer [25]. However, thermistors in direct contact with the TM are not practical for everyday use.

Instead of being in direct contact with the TM, today’s tympanic thermometers measure the thermal radiation emitted from the TM and the ear canal, and have therefore been called infrared radiation emission detectors (IRED). Because the amount of thermal radiation emitted is in proportion to the membrane’s temperature, IRED accurately estimates TM temperature [16]. In contrast with other sites of temperature measurement, the TM’s blood supply is very similar in temperature and location to the blood bathing the hypothalamus, the site of the body’s thermoregulatory centre. It is, therefore, an ideal location for core temperature estimation [26][27]. Crying, otitis media or earwax have not been shown to change tympanic readings significantly.

Much has been published both in support of [15][16][26][27] and against [29]-[32] the use of infrared tympanic thermometers in clinical practice, and it is no wonder that many physicians remain skeptical about measurement reliability. Results of a questionnaire completed by randomly selected members of the American Academy of Pediatrics and the American Academy of Family Physicians demonstrated that 78% of respondents had used infrared thermometers at least once; 65% of paediatricians and 64% of family practitioners were current users [32]. The most commonly reported causes for the discontinued use of tympanic thermometers were inaccuracy or lack of staff trust with the device.

Most studies that compare the accuracy of tympanic thermometers with other classical measures of body temperature evaluate the accuracy of tympanic readings by comparing them with rectal, oral or axillary measurements. Given the variations of temperature ranges with each of these methods and the limitations of their accuracy discussed above, using any one method as a ‘benchmark’ or ‘gold standard’ is misleading. Because estimates of core temperature measured at different body sites will vary, an effort has been made by manufacturers of IREDs to correlate tympanic readings to rectal or oral equivalents [16].

These conversion scales (known as ‘offsets’) convert the measured ear temperature to one that would be found at a different site, allowing a user to define more easily a fever from a measurement in the ear. The offsets are based on an algorithm that transforms a subject’s tympanic temperature to that found at either the oral or rectal site. However, the data used to develop these offsets may not be readily applicable to the paediatric population. Some researchers advise eliminating these adjusted modes and simply using unadjusted ear temperature (Table 1) [16][18][21][43].

TABLE 1
Normal temperature ranges
Measurement methodNormal temperature range
Rectal36.6°C to 38°C (97.9°F to 100.4°F)
Ear35.8°C to 38°C (96.4°F to 100.4°F)
Oral35.5°C to 37.5°C (95.9°F to 99.5°F)
Axillary36.5°C to 37.5°C (97.8°F to 99.5°F)

Factors related to the patient, instrument, technique and environment contribute to the variability of ear-based temperature measurements. For example, the ear canal’s structure, probe design and probe positioning affect how well the canal is sealed from ambient influences and what parts of the tympanic membrane, ear canal wall, and perhaps skin surface, are in the thermometers field of view [33]. To get an accurate reading of tympanic temperature, the infrared probe (up to 8 mm in diameter) must be small enough to be deeply inserted into the meatus to allow orientation of the sensor against the TM [28]. While this is of less concern in children older than two years of age whose meatus is wider than 8 mm, the average diameter of the meatus in young children (4 mm at birth, 5 mm at two years of age) can cause complications for tympanic thermometry. When the probe is too large, it will detect infrared emissions from both the TM and the proximal meatus wall. Because the thermometer averages the two surface temperatures, it can produce an erroneously low reading. It is generally recommended that a slight tug of the pinna to straighten the ear canal can improve accuracy and consistency.

Also, each different brand of ear thermometers has its own design, technology, offsets and operating instructions that affect its reliability, accuracy and use. Consumer and professional units are available; the latter are designed to be more durable to withstand day-to-day use in a professional setting. While many current brands exist, the reliability of different instruments seems to be comparable, if the manufacturer’s instructions are followed properly.

The authors of a recent systematic review and meta-analysis comparing infrared tympanic and rectal thermometry concluded that infrared tympanic thermometery lacks sufficient agreement with their defined gold standard—rectal thermometry—to be used with confidence in situations where body temperature needs to be measured with precision [35].

Other devices

Infrared arterial temperature can be measured with a device that is passed over the front of the forehead to the temporal area. This relatively new method of body temperature measurement has been shown to be more accurate than tympanic thermometry and better tolerated than rectal thermometry [36].

In a recent trial in a paediatric emergency department, the temporal artery device was shown to have a sensitivity of about 80% to identify fever (as determined by rectal measurements). When a cut-off temperature over 37.7ºC was used on the temporal artery device to define fever, the sensitivity improved to 90% for identifying a fever of >38ºC as measured by the rectal thermometer, but the specificity dropped to about 50%. Moreover, parents’ use of a similar device resulted in inadequate agreement with rectal temperatures [37]. Other studies have concluded that temporal artery thermometers (TATs) may be an effective screening tool for homes and offices, but not in clinical settings where investigations and treatment may be affected by limited sensitivity in infants younger than 3 years of age[38][39][40]. TATs do show promise in situations where rectal thermometry is contraindicated (eg, during a surgical operation on the rectum or perineum, or when an inflammatory disorder of the rectum, congenital ano-rectal malformation, fistulae or fever/neutropenia are present). TATs are also susceptible in a setting where radiant warmers are in use [39] [44].

Temporal artery thermometry, though promising, may be a more useful tool for screening children at low risk in an acute care setting; this method cannot yet be recommended for home or hospital use when definitive measurements are required. Further studies involving large numbers of subjects are required before any alternative method replaces rectal or axillary/oral thermometry as “the gold standard” in children younger than 3 years of age. Current pricing for temporal artery temperature devices is also significantly higher than for oral/rectal thermometers, which may be a complicating factor for home use or for institutions where many devices are required.

Pacifier thermometers may not be feasible for sick infants (particularly if in respiratory distress) and have been shown to be less accurate than rectal thermometers [41].

Conclusion

While it is evident that all devices available currently to measure temperature in children have their strengths and weaknesses, the choice made by parents may be influenced by the convenience of use, cost and advertising. For professionals, the older, time-honoured methods may be chosen because they are deeply entrenched in the medical literature and there is no groundswell for change. However, in keeping with environmental concerns, mercury thermometers should no longer be used.

Based on the evidence currently available, the relative ease, speed, accuracy and safety of the infrared tympanic thermometer warrant its inclusion in the group of currently available instruments for temperature measurement in older children. Nonetheless, children who are younger than two years of age should continue to have their temperature taken rectally until an adequate probe for tympanic thermometry is designed or until other methods of measurement are shown to be accurate and reliable in larger studies (Table 2).

TABLE 2
Summary of recommended temperature measurement techniques
Canadian Task Force on Preventive Health - Strength of Recommendations: B, II [42]
AgeRecommended technique
Birth to 2 years1. Rectal (definitive)
2. Axillary (screening low risk children)
Over 2 years to 5 years1. Rectal (definitive)
2. Axillary, Tympanic (or Temporal Artery if in hospital) (screening)
Older than 5 years1. Oral (definitive)
2. Axillary, Tympanic (or Temporal Artery if in hospital) (screening)

Acknowledgements

The committee thanks Dr. Howard Levitt for this research in the preparation of the original version of this position statement.

COMMUNITY PAEDIATRICS COMMITTEE
Members: Cecilia Baxter MD (1998-2004), Mark Feldman MD, William James MD (2002-2004), Mia Lang MD, Denis Leduc MD (Chair, 1998-2004), Cheryl Mutch MD, Michelle Ponti MD (chair), Linda Spigelblatt MD, Sandra Woods MD (1998-2004), David Wong MD (Board representative)
Liaison: Richard Haber MD, CPS Community Paediatrics Section
Principal authors: Denis Leduc MD, Sandra Woods MD
Revision: Mark Feldman MD, Larry Pancer MD

References

  1. Mackowiak PA, Wasserman SS, Levine MM. A critical appraisal of 98.6 degrees F, the upper limit of the normal body temperature, and other legacies of Carl Reinhold August Wunderlich. JAMA 1992;268:1578-80.
  2. Herzog LW, Coyne LJ. What is fever? Normal temperature in infants less than 3 months old. Clin Pediatr 1993;32:142-6.
  3. Teach SJ, Fleisher GR. Duration of fever and its relationship to bacteremia in febrile outpatients three to 36 months old. The Occult Bacteremia Study Group. Pediatr Emerg Care 1997;13:317-9.
  4. Grossman M. Management of the febrile patient. Pediatr Infect Dis. 1986;5:730-4.
  5. McCarthy PL. The Evaluation and Management of Febrile Children. New York : Appleton-Century-Crofts,1988.
  6. Soman M. Diagnostic workup of febrile children under 24 months of age: A clinical review. West J Med 1982;137:1-12.
  7. Baskin MN . The prevalence of serious bacterial infections by age in febrile infants during the first 3 months of life. Pediatr Ann 1993;22:462-6.
  8. Brik R, Hamissah R, Shehada N, et al. Evaluation of febrile infants under 3 months of age: Is routine lumber puncture warranted? Isr J Med Sci 1997;33:93-7.
  9. Kleiman MB . The complaint of persistent fever. Recognition and management of pseudo fever of unknown origin. Pediatr Clin North Am 1982;29:201-8.
  10. McClung HJ. Prolonged fever of unknown origin in children. Am J Dis Child 1972;124:544-50.
  11. Pizzo PA, Lovejoy FH, Smith DH. Prolonged fever in Children: Review of 100 Cases. Pediatrics 1975;55:468-73.
  12. Lieu TA, Baskin MN, Schwartz JS, Fleisher GR. Clinical and cost effectiveness of outpatient strategies for management of febrile infants. Pediatrics 1992;89:1135-44.
  13. McCarthy PL. Fever. Pediatr Rev 1998;19:401-7.
  14. Brown PJ, Christmas BF, Ford RP. Taking an infant’s temperature: Axillary or rectal thermometer? N Z Med J 1992;105:309-11.
  15. Romano MJ, Fortenberry JD, Autrey E, et al. Infrared tympanic thermometry in the pediatric intensive care unit. Crit Care Med 1993;21:1181-5.
  16. Chamberlain JM, Terndrup TE, Alexander DT, et al. Determination of normal ear temperature with an infrared emisssion detection thermometer. Ann Emerg Med 1995;25:15-20.
  17. Robinson JL, Seal RF, Spady DW, Joffres MR. Comparison of esophageal, rectal, axillary, bladder, tympanic, and pulmonary artery temperatures in children. J Pediatr 1998;133:553-6.
  18. Erickson RS, Woo TM. Accuracy of infrared thermometry and traditional temperature methods in young children. Heart Lung 1994;23:181-95.
  19. Blainey CG. Site selection in taking body temperature. Am J Nurs 1974;74:1859-61.Am J Nurs 1974;74:1859-61.
  20. Kenney RD , Fortenberry JD, Surratt SS, Ribbeck BM, Thomas WJ. Evaluation of an infrared tympanic membrane thermometer in pediatric patients. Pediatrics 1990;85:854-8.
  21. Jaffe DM. What’s hot and what’s not: The gold standard for thermometry in emergency medicine. Ann Emerg Med 1995;25:97-9.
  22. Kresch MJ. Axillary temperature as a screening test for fever in children. J Paediatr 1984;104:596-9.
  23. Morley C, Hewson P, Thornton A, Cole T. Axillary and rectal temperature measurements in infants. Arch Dis Child 1992;67(1):122-5.
  24. Press S, Quinn B. The pacifier thermometer: Comparison of supralingual with rectal temperatures in infants and young children. Arch Pediatr Adolesc Med 1997;151:551-4.
  25. Benzinger M, Benzinger T. Tympanic clinical temperature. In: Thomas HP, Murray TP, Shepard RL, eds. Fifth Symposium on Temperature. Washington : American Institute of Physics, Instrument Society of America , National Bureau of Standards 1972:2089-2102.
  26. Terndrup T, Crofton D, Mortelliti A, Kelley R, Rajk J. Estimation of contact tympanic membrane temperature with a noncontact infrared thermometer. Ann Emerg Med 1997;30:171-5.
  27. Childs C, Harrison R, Hodkinson C. Tympanic membrane temperature as a measure of core temperature. Arch Dis Child 1999;80:262-6.
  28. Smith J. Are electronic thermometry techniques suitable alternatives to traditional mercury in glass thermometry techniques in the paediatric setting? Journal of Advanced Nursing 1998;28(5):1030-9.
  29. Romanovsky A, Quint P, Benikova Y, Kiesow L. A difference of 5 degrees C between ear and rectal temperatures in a febrile patient. Am J Emerg Med 1998;125:83-5.
  30. Petersen-Smith A, Barber N, Coody D, West M, Yetman R. Comparison of aural infrared with traditional rectal temperatures in children from birth to age three years. J Pediatr 1994;125:83-5.
  31. Petersen M, Hauge H. Can training improve the results with infrared tympanic thermometers? Acta Anaesthesiol Scand 1997;41:1066-70.
  32. Modell J, Katholi C, Kumaramangalam S, Hudson E, Graham D. Unreliability of the infrared tympanic thermometer in clinical practice: a comparative study with oral mercury and oral electronic thermometers. South Med J 1998;91:649-54.
  33. Silverman B, Daley W, Rubin J. The use of infrared ear thermometers in pediatric and family practice offices. Public Health Rep 1998;113:268-7.
  34. Benzinger M. Tympanic thermometry in surgery and anaesthesia. JAMA 1969;209:1207-11.
  35. Craig J, Lancaster G, Taylor S, Williamson P, Smyth R. Infrared ear thermometry compared with rectal thermometry in children: a systematic review. The Lancet 2002;360(9333):603.
  36. Greenes D, Fleisher G. Accuracy of a Noninvasive Temporal Artery Thermometer for Use in Infants. Arch Pediatr Adolesc Med 2001;155:376.
  37. Schuh S, Komar L, Stephens D, Chu L, Read S, Allen U. Comparison of the temporal artery and rectal thermometry in children in the emergency department. Pediatr Emerg Care 2004;20(11):736-41.
  38. Siberry GK, Diener-West M, Schappell E,Karron RA. Comparison of temple temperatures with rectal temperatures in children under two years of age. Clin Pediatr (Phila) 2002; 41(6):405-14.
  39. Hoffman RJ, Etwaru K, Dreisinger N, Khokhar A, Husk G. Comparison of temporal artery thermometry and rectal thermometry in febrile pediatric emergency department patients. Pediatr Emerg Care 2013;29(3):301-4.
  40. Teller J, Ragazzi M, Simonetti GD, Lava SA. Accuracy of tympanic and forehead thermometers in private paediatric practice. Acta Paediatr 2014;103(2):e80-3.
  41. Callanan D. Detecting fever in young infants: reliability of perceived, pacifier, and temporal artery temperatures in infants younger than 3 months of age. Pediatr Emerg Care 2003;19(4):240-3.
  42. Canadian Task Force on Preventive Health Care. New grades for recommendations from the Canadian Task Force on Preventive Health Care. CMAJ 2003;169:207-8.
  43. Takayama,JI, TengW, Uyemoto J, Newman TB, Pantell RH. Body temperature of newborns: What is normal? Clin Pediatr (Phila) 2000;39(9):503-10.
  44. Reynolds M, Bonham L, Gueck M, et al. Are temporal artery temperatures accurate enough to replace rectal temperature measurement in pediatric ED patients? J Emerg Nurs 2014;40(1):46-50.

Disclaimer: The recommendations in this position statement do not indicate an exclusive course of treatment or procedure to be followed. Variations, taking into account individual circumstances, may be appropriate. Internet addresses are current at time of publication.